Loading...

Syphilis EIA Total Antibody

One step sandwich assay

Syphilis EIA Total antibody

Principle of the test

NewBio-Syphilis TA is a one step sandwich assay using recombinant antigens for enhanced sensitivity and specificity. The assay is for use with plasma and serum samples and will detect antibodies at all stages of infection. Antibodies are captured by recombinant antigens on the plate and marked by HRP recombinant conjugates for visualisation with TMB substrate. The reagents and protocol ensure ease of use and assay control.

Specificity

A study on 300 donor serum showed 100% specificity. (95% confidence limits 99.7-100%)

A study on 300 donor EDTA plasma showed 100% specificity. (95% confidence limits 99.7-100%)

Sensitivity

A study on 100 syphilis positive samples showed 100% sensitivity. (95% confidence limits 99.6-100%)

Analytical Sensitivity

NewBio-Syphilis TA has a sensitivity of 0.0015 IU/ml against the 1st IS for human syphilitic plasma IgG and IgM NIBSC code: 05/132

Key Features

  • Quality Assured - 18 years experience

  • Ready to use stable liquid reagents

  • 24 months shelf life

  • Qualitative and semi-quantitative protocols

  • Results ready in 60 minutes

  • Automated reading

  • Suitable for automation

  • Convenient kit sizes

NewBio Syphilis EIA Total antibody Test Kit
NewBio Syphilis EIA 96 Total antibody Test Kit
  • Product No:
    NB010
  • Quantity:
    96 Tests
  • Stability:
    24 months @ 2-8°C
NewBio Syphilis EIA Total antibody 96 tests
  • One step sandwich assay for Syphilis IgG, IgM and IgA
  • Recombinant antigens 15kDa, 17kDa and 47kDa
  • All reagents ready to use. (except 20 x wash buffer)
  • Positive Control
    Negative Control
    Solid phase 8 well strips
    HRP Conjugate
    TMB Substrate
    Stop solution
    Wash buffer 20 x concentration
  • Simple protocol: 2 x 30 minute incubations
    Manual
    Automatable
  • Standard 50µl volumes samples and reagents
  • Sample and reagent check
    Single visual/readable colour change (OD550) on addition of conjugate to sample
NewBio Syphilis EIA 480 Total antibody Test Kit
  • Product No:
    NB011
  • Quantity:
    480 Tests
  • Stability:
    24 months @ 2-8°C
NewBio Syphilis EIA Total antibody 480 tests
  • One step sandwich assay for Syphilis IgG, IgM and IgA
  • Recombinant antigens 15kDa, 17kDa and 47kDa
  • All reagents ready to use. (except 20 x wash buffer)
  • Positive Control
    Negative Control
    Solid phase 8 well strips
    HRP Conjugate
    TMB Substrate
    Stop solution
    Wash buffer 20 x concentration
  • Simple protocol: 2 x 30 minute incubations
    Manual
    Automatable
  • Standard 50µl volumes samples and reagents
  • Sample and reagent check
    Single visual/readable colour change (OD550) on addition of conjugate to sample